| dc.contributor.advisor | Ambrósio Junior, Renato | |
| dc.contributor.author | Reis, Eric | |
| dc.date.accessioned | 2026-10-13T19:18:03Z | |
| dc.date.available | 2026-10-13T19:18:03Z | |
| dc.date.issued | 2026-03-10 | |
| dc.identifier.citation | REIS, Eric. Avaliação da repetibilidade e intercambiabilidade de um tomógrafo de segmento anterior com tecnologia Swept Source (CASIA II). 2025. 51 f. Dissertação (Mestrado em Medicina) – Universidade Federal do Estado do Rio de Janeiro, Rio de Janeiro, 2025. | pt_BR |
| dc.identifier.uri | http://hdl.handle.net/unirio/15453 | |
| dc.description.sponsorship | n/a | pt_BR |
| dc.language.iso | Portuguese | pt_BR |
| dc.rights | openAccess | pt_BR |
| dc.title | Avaliação da repetibilidade e intercambiabilidade de um tomógrafo de segmento anterior com tecnologia Swept Source (CASIA II) | pt_BR |
| dc.title.alternative | Repeatability and interchangeability assessment of an anterior segment tomographer with Swept Source technology (CASIA II) | pt_BR |
| dc.type | masterThesis | pt_BR |
| dc.contributor.referee | Ambrósio Junior, Renato | |
| dc.contributor.referee | Motta, Mario Martins dos Santos | |
| dc.contributor.referee | Carvalho, Marcella Quaresma Salomão Hoyer de | |
| dc.contributor.referee | Luz, Allan Cezar de Sousa | |
| dc.degree.department | CCBS | pt_BR |
| dc.degree.grantor | Universidade Federal do Estado do Rio de Janeiro - UNIRIO | pt_BR |
| dc.degree.level | Mestrado Profissional | pt_BR |
| dc.degree.local | Rio de Janeiro, RJ | pt_BR |
| dc.degree.program | Programa de Pós-Graduação em Medicina | pt_BR |
| dc.subject.cnpq | MEDICINA | pt_BR |
| dc.subject.cnpq | CIÊNCIAS DA SAÚDE | pt_BR |
| dc.subject.en | Keratoconus | pt_BR |
| dc.subject.en | Repeatability | pt_BR |
| dc.subject.en | Refractive surgery | pt_BR |
| dc.subject.en | Anterior segment tomography | pt_BR |
| dc.description.abstracten | This study aimed to evaluate the repeatability and interchangeability of anterior segment parameters obtained by the CASIA II Swept Source Tomographer in three groups: normal, keratoconus, and post-refractive surgery. This is a prospective and cross-sectional study including 44 eyes from 44 individuals. Each participant obtained five consecutive measurements. Each measurement evaluated 14 parameters. For interchangeability analysis, total and epithelial pachymetry were compared with the Spectral Domain RtVue Tomographer. Repeatability was assessed using the intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), coefficient of variation (CV), and coefficient of repeatability (CR). All exams considered central, paracentral, and peripheral zones according to the CASIA II sectoral layout. Normal corneas showed the highest percentage of sectors with excellent ICC, followed by post-refractive surgery and keratoconus. Pachymetry was the parameter with the greatest number of significant differences among groups, especially in peripheral regions. All parameters showed mean ICC > 0.80, with better performance in central and paracentral regions. Sw, CV, and CR showed favorable results in all groups, although with higher values in peripheral regions and in corneas with keratoconus. Agreement between CASIA II and RtVue was good for total and epithelial pachymetry in central and paracentral regions. In the keratoconus group, the thinnest point showed higher repeatability indices compared to the other groups. It is concluded that CASIA II demonstrates high repeatability and good interchangeability with RtVue, even in irregular corneal conditions, reinforcing its clinical applicability in detailed corneal evaluation. | pt_BR |
| dc.degree.country | Brasil | pt_BR |
| dc.description.sponsordocumentnumber | n/a | pt_BR |
| dc.description.abstractpt | Este estudo teve como objetivo avaliar a repetibilidade e a intercambiabilidade dos parâmetros do segmento anterior obtidos pelo tomógrafo CASIA II, com tecnologia Swept Source, em córneas normais, com ceratocone e pós-cirurgia refrativa. Trata-se de um estudo prospectivo e transversal envolvendo 44 olhos de 44 indivíduos, divididos entre os três grupos. Cada participante foi submetido a cinco medidas consecutivas de 14 parâmetros do CASIA II. Para análise de intercambiabilidade, a paquimetria total e epitelial foram comparadas com o tomógrafo Spectral Domain RtVue. A repetibilidade foi avaliada por meio do coeficiente de correlação intraclasse (ICC), desvio padrão intraindivíduo (Sw), coeficiente de variação (CV) e coeficiente de repetibilidade (CR), considerando zonas central, paracentral e periférica, conforme layout setorial do CASIA II. As córneas normais apresentaram maior percentual de setores com ICC excelente, seguidas por pós-LVC e ceratocone. A paquimetria foi o parâmetro com maior número de diferenças significativas entre os grupos, especialmente em regiões periféricas. Todos os parâmetros apresentaram ICC médio > 0,80, com melhor desempenho nas regiões centrais e paracentrais. Sw, CV e CR mostraram resultados favoráveis em todos os grupos, embora com valores mais elevados nas regiões periféricas e em córneas com ceratocone. A concordância entre CASIA II e RtVue foi boa para paquimetria total e epitelial nas regiões centrais e paracentrais. No grupo ceratocone, o ponto mais fino apresentou índices de repetibilidade superior aos demais grupos. Conclui-se que o CASIA II apresenta alta repetibilidade e boa intercambiabilidade com o RtVue, mesmo em condições corneanas irregulares, reforçando sua aplicabilidade clínica na avaliação detalhada da córnea. | pt_BR |
| dc.subject.pt | Ceratocone | pt_BR |
| dc.subject.pt | Repetibilidade | pt_BR |
| dc.subject.pt | Cirurgia refrativa | pt_BR |
| dc.subject.pt | Tomografia segmento anterio | pt_BR |